Log in

Table II Long-term ECG signal correlation and average peak-to-peak amplitude

From: Rapid Fabrication Technique for Dry Electrocardiography Electrodes Using Carbon Nanotube/Polydimethylsiloxane Composite

 

Day 0

Day 1

Day 2

Day 3

Day4

CNT/PDMS-based electrode and reference electrode ECG signal

Correlation

0.9732

0.9755

0.9711

0.9721

0.9701

Ag/AgCl electrode and reference electrode ECG signal

Correlation

0.9983

0.9935

0.9630

0.9432

0.9325

Average peak-to-peak amplitude of ECG signal

Reference Ag/AgCl electrode (A)

0.70

0.70

0.68

0.68

0.65

CNT/PDMS-based electrode (B)

0.71

0.69

0.69

0.69

0.64

Ag/AgCl electrode (C)

0.70

0.70

0.61

0.39

0.32

Ratio of (B) to (A)

1.01

0.98

1.01

1.01

0.98

Ratio of (C) to (A)

1.0

1.0

0.89

0.57

0.49